Axon radius estimation with Oscillating Gradient Spin Echo (OGSE) diffusion MRI
DOI:
https://doi.org/10.62721/diffusion-fundamentals.18.665Keywords:
axon diameter, diffusion MR, oscillating gradient spin echo, optimised gradient waveform, ActiveAx.Abstract
The estimation of axon radius provides insights into brain function [1] and could provide progression and classification biomarkers for a number of white matter diseases [2-4]. A recent in silico study [5] has shown that optimised gradient waveforms (GEN) and oscillating gradient waveform spin echo (OGSE) have increased sensitivity to small axon radius compared to pulsed gradient spin echo (PGSE) diffusion MR sequences. In a follow-up study [6], experiments with glass capillaries show the practical feasibility of GEN sequences and verify improved pore-size estimates. Here, we compare PGSE with sine, sine with arbitrary phase, and square wave OGSE (SNOGSE, SPOGSE, SWOGSE, respectively) for axon radius mapping in the corpus callosum of a rat, ex-vivo. Our results suggest improvements in pore size estimates from OGSE over PGSE, with greatest improvement from SWOGSE, supporting theoretical results from [5] and other studies [7-9].Downloads
Published
2013-07-01
How to Cite
Siow, B., Drobnjak, I., Ianus, A., Christie, I. N., Lythgoe, M. F., & Alexander, D. C. (2013). Axon radius estimation with Oscillating Gradient Spin Echo (OGSE) diffusion MRI. Diffusion Fundamentals, 18. https://doi.org/10.62721/diffusion-fundamentals.18.665